Solution Turn on the Quantum Blue Reader and check the external power supply unit.
KESO AccessPoint-(blue) Reader 13.56MHz.
Problem The Quantum Blue Reader is blocked
The Quantum Blue Reader is now ready to be used with your language of choice.
Problem The Quantum Blue Reader is entirley blocked
The Quantum Blue Reader must only be operated by qualified personnel who have been appropriately trained.
Improper use of the Quantum Blue Reader may cause personal injuries
This Quantum Blue Reader contains an internal RFID reader therefore, default setting is“Rfid_intern”.
The Quantum Blue Reader should only be used with specified tests by trained
for 2 seconds to turn on the Quantum Blue Reader see chapter 5.2.
Cause An error during the initialization or during the operation of the Quantum Blue Reader may have occurred.
Servicing of the Quantum Blue Reader must only be performed by Field Service Engineers trained by BÜHLMANN Laboratories AG.
Cause The RFID Chip Card was not placed long enough at position“A” on the Quantum Blue Reader.
Please note that only ONE language can be displayed at a time on the Firmware of the Quantum Blue Reader.
Take out the Quantum Blue Reader from its protective transportation case
The instructions should provide the user with the necessary information for proper maintenance and care of the Quantum Blue Reader.
by the firmware of the Quantum Blue Reader.
Solution Hold RFID Chip Card in the correct position on the Quantum Blue Reader(see chapter 3.1,“A”)
hold the respective RFID Chip Card for a few seconds at position“A” on the Quantum Blue Reader(see also chapter 3.1)
Problem The Quantum Blue Reader displays an“INVALID” test result.